Instructions for Use
75 Pages
Preview
Page 1
Table of Contents 1.
2.
3.
4.
5.
6.
INTRODUCTION... 1 1.1 SCOPE ... 1 1.2 FOREWORD ... 1 1.3 APPLICABLE EUROPEAN DIRECTIVES/NORMS ... 1 1.4 INTENDED USE ... 1 1.5 SYMBOLS DISPLAYED ON THE STERILIZER ... 2 1.6 GENERAL AND SAFETY RECOMMENDATIONS ... 2 UNPACKING ... 3 2.1 UNPACKING THE STERILIZER ... 3 2.2 STANDARD ACCESSORIES ... 3 UNIT DESCRIPTION ... 4 3.1 FRONT VIEW ... 4 3.2 SERVICE DOOR ... 4 3.3 REAR VIEW ... 5 3.4 DESCRIPTION OF THE INTERNAL WATER TANKS ... 6 3.5 CHAMBER RACK ... 7 3.6 USABLE SPACE IN THE CHAMBER ... 7 INSTALLATION ... 8 4.1 SETUP ... 8 4.1.1 Securing the sterilizer with a safety bracket ... 8 4.2 ELECTRICAL POWER SUPPLY ... 9 4.3 PRINTER (optional) ... 9 4.4 LISAWARE - CONNECTING THE STERILIZER TO A COMPUTER (optional) ... 10 4.4.1 LAN connection ... 10 4.4.2 Serial connection ... 10 4.5 DEMINERALIZER (optional) ... 11 4.5.1 Connecting an external water supply system (demineralizer) ... 11 4.6 CONTINUOUS DRAINING (optional) ... 11 4.6.1 Connecting the drain tube ... 11 GETTING STARTED ... 12 5.1 THE USER INTERFACE ... 12 5.2 INITIAL WARNINGS AND SLEEP MODE ... 13 5.3 DATE-CLOCK SETTING ... 13 5.4 FILLING THE CLEAN WATER TANK ... 14 5.4.1 Manual filling ... 14 5.4.2 Automated water supply (optional) ... 14 5.5 DRAINING THE USED WATER TANK ... 15 5.5.1 Manual draining ... 15 5.5.2 Continuous draining ... 15 5.6 MEMORY CARD ... 15 5.6.1 Inserting / removing the memory card ... 15 PROGRAMMING ... 16 6.1 SETUP MENU ... 16 6.1.1 Language ... 16 6.1.2 Sleep mode... 17 6.1.3 Printer ... 17 6.1.4 Label printer (available only when a LisaSafe label printer is connected) ... 17 6.1.5 Automatic printing (available only when a LisaSafe label printer is connected) ... 17 6.1.6 Manual printing (available only when a LisaSafe label printer is connected) ... 17 6.1.7 Storage time/weeks (available only when a LisaSafe label printer is connected) ... 17 6.1.8 User name ... 17 6.1.9 Date-Clock setting ... 17 6.1.10 Date format ... 18 6.1.11 Clock format... 18 6.1.12 Display contrast ... 18 6.1.13 Display backlight ... 18 6.1.14 Acoustic tones ... 18 6.1.15 Night cycle shortcut ... 19 6.1.16 Connection to PC/Log ... 20 6.1.17 RSDATA port selection ... 20 6.2 TEST CYCLES ... 21 6.3 DELAYED CYCLE START ... 22 6.4 CYCLE HISTORY ... 23 6.5 MAINTENANCE ... 24 6.6 SERVICE ... 24 6.7 UTILITY ... 24
a
6.7.1 System info ... 25 6.7.2 Formatting the memory card ... 25 6.8 USER AUTHENTICATION ... 25 6.9 FUNCTIONS AVAILABLE FOR THE ADMINISTRATOR ... 26 6.9.1 Access the administrator functions ... 26 6.9.2 Modifying a user name ... 27 6.9.3 Changing the administrator’s password ... 28 6.9.4 Resetting a user’s password to the default value (0000) ... 29 6.9.5 Deleting a user... 30 6.9.6 Adding a new user ... 31 6.9.7 Setting the user identification options for all users ... 32 6.10 FUNCTIONS AVAILABLE TO THE OPERATOR(S): PASSWORD MODIFICATION ... 33 6.11 STARTING AND ENDING A STERILIZATION CYCLE WITH IDENTIFICATION OF THE USER ... 35 6.11.1 Starting a sterilization cycle ... 35 6.11.2 End of a sterilization cycle ... 36 7. RUNNING A STERILIZATION CYCLE ... 37 7.1 THE AVAILABLE STERILIZATION CYCLES ... 37 7.1.1 Starting a sterilization cycle ... 39 7.1.2 ECO-Dry feature ... 39 7.1.3 Cycle in-progress ... 40 7.1.4 End of cycle ... 40 7.2 MANUAL STOP ... 41 7.3 REAL-TIME CYCLE DATA INFORMATION ... 42 7.4 CYCLE DATA SUMMARY ... 43 8. DISPLAY MESSAGES ... 44 9. ALARMS ... 46 10. ALARM CODE TABLE ... 47 11. MAINTENANCE ... 49 11.1 MAINTENANCE PROGRAM ... 49 11.2 CLEANING THE DOOR SEAL ... 50 11.3 CLEANING THE CHAMBER AND CHAMBER COMPONENTS ... 50 11.4 CLEANING THE CHAMBER FILTER ... 50 11.5 CLEANING THE EXTERNAL STERILIZER SURFACES ... 51 11.6 REPLACING THE BACTERIOLOGICAL FILTER ... 51 11.7 REPLACING THE DUST FILTER ... 51 11.8 CLEANING THE WATER TANKS ... 52 11.9 REPLACING THE DOOR SEAL ... 53 11.10 SERVICE CONDUCTED BY AN AUTHORIZED SERVICE TECHNICIAN ... 54 12. USE OF THE MEMORY CARD ... 55 12.1 TECHNICAL CHARACTERISTICS OF THE MEMORY CARD ... 55 12.2 READING OF MEMORY CARD DATA WITH A PC/MAC ... 55 12.3 MINIMUM HARDWARE REQUIREMENTS FOR A PC/MAC ... 55 12.4 CONNECTING THE EXTERNAL USB CARD READER TO YOUR PC/MAC ... 56 12.5 SAVED FILE ... 56 12.6 CONTROL CODE ... 56 12.7 FILE NAMES ... 56 12.8 DIRECTORY NAME ... 56 12.9 MEMORY CARD MANAGEMENT ... 57 12.10 SAVING A FILE ... 58 13. TROUBLESHOOTING ... 59 14. RECYCLING / DISPOSAL INSTRUCTIONS ... 60 Appendix 1 TECHNICAL CHARACTERISTICS ... 61 Appendix 2 STERILIZATION LOAD PREPARATION ... 62 Appendix 3 MAINTENANCE OF DENTAL HANDPIECES ... 63 Appendix 4 BOWIE AND DICK TEST ... 64 Appendix 5 HELIX TEST ... 65 Appendix 6 VACUUM TEST ... 66 Appendix 7 WATER QUALITY ... 67 Appendix 8 EXAMPLE OF A CYCLE DATA REPORT ... 68 Appendix 9 ACCESSORIES ... 69 Appendix 10 HELIX TEST DOCUMENTATION FORM ... 71
b
1.
INTRODUCTION
1.1
SCOPE
The purpose of this user manual is to provide you with information about Lisa 517/522 sterilizers to ensure: proper installation and set-up. optimal use. safe and reliable operation. compliance with regular maintenance and servicing requirements. NOTE:
1.2
All drawings, images and texts contained in this manual are the property of the manufacturer. All rights reserved. Even partial duplication of drawings, images or text is prohibited. The information contained in this document is subject to change without notice.
FOREWORD
There are two types of users who may operate the sterilizer and who shall read this manual carefully: Administrator The head of the clinic / practice, who is legally responsible for the efficiency of the hygiene protocol in place as well as the sterilization process. He/she is also responsible for the OPERATORS’ training and the correct operation and maintenance of the equipment. Operator The person(s) who use(s) the sterilizer according to the ADMINISTRATOR’S instructions.
1.3
APPLICABLE EUROPEAN DIRECTIVES/NORMS
Lisa 517/522 sterilizers conform to the following European Directives: Medical Device Directive 93/42/CEE for devices class IIb, in accordance with rule 15 - Appendix IX of the above directive. Directive 97/23/CEE (Pressure Equipment Directive – PED) for every sterilization chamber designed and manufactured in conformity to Annex I and to the procedure described in form D1 annex III. Directive 2002/96/CEE (RAEE) for disposal of parts coming from electrical or electronic parts. The sterilizer has been developed, produced and tested in accordance with the European Norm relative to small water steam sterilizers EN13060, and with the applicable safety norms (see Appendix 1). In the enclosed documents, you will find the Declaration of Conformity and a Warranty Card specific to your sterilizer.
1.4
INTENDED USE
Lisa 517/522 sterilizers are fully automated bench top steam sterilizers that generate steam using electrical heaters. Lisa 517/522 sterilizers are used for medical purposes, e.g. in general medical practices, dental offices, facilities for personal hygiene and beauty care and veterinary practices. They are also used for materials and equipment that is likely to be exposed to blood or body fluids, e.g. instruments used by beauty therapists, tattooists, body piercers and hairdressers. The types of loads that can be sterilized with Lisa sterilizers are described in Table 1 of the reference technical norm EN 13060. These loads include solid, porous, hollow loads type A and hollow loads type B, unwrapped, single wrapped and double wrapped. Lisa sterilizers cannot be used to sterilize liquids or pharmaceutical products.
1
1.5
SYMBOLS DISPLAYED ON THE STERILIZER
Consult this table whenever one you see one of these symbols in this manual or on the sterilizer. ATTENTION Where this symbol is displayed on the sterilizer, the user must refer to this document. When shown in the user manual this symbol means ATTENTION IMPORTANT NOTES. To disregard the instructions given in this manual, incorrect use, poor maintenance or servicing by unauthorized personnel, clears the manufacturer of any responsibility for warranty and any other claims. HOT SURFACES This symbol is displayed at the front of the sterilizer beneath the chamber door. It reminds the user to take special care to avoid burns when dealing with the sterilization load, the sterilization chamber, the chamber door and the area around the chamber door.
The material the sterilizer is composed of must be disposed according to the directive 2002/96/CEE.
1.6
GENERAL AND SAFETY RECOMMENDATIONS o
o o o o o o o o o o
The user is responsible for the proper installation, the correct use and maintenance of the sterilizer in accordance with the instructions listed in this manual. For further information call your local service provider. The sterilizer has not been designed for the sterilization of liquids. The sterilizer must not be used in the presence of gas or explosive vapours. The chamber is automatically heating to 120°C as soon as the sterilizer is switched on. The trays and the sterilization load will be hot at the end of each cycle. Use tray or cassette holders to empty the sterilization chamber. Do not exceed the maximum load weight limits as specified in this manual (see § 7.1) Do not remove the name plate or any label from the sterilizer. To avoid electrical short circuits, do not pour water or any other liquid over the sterilizer. Switch off the sterilizer and unplug the mains cable before inspecting, carrying out maintenance or servicing the sterilizer. Repairs, maintenance or service must be carried out by authorized W&H service technicians only with the use of original spare parts. In case of transport:
- Drain both water tanks completely (§ 5.5). - Allow the sterilization chamber to cool down. - Use original or appropriate packaging. Not observing the instructions as specified in this manual can lead to unsafe operation for the user.
2
2.
UNPACKING
2.1
UNPACKING THE STERILIZER If the sterilizer had been kept in a place with temperature and humidity different from the installation location, wait for an appropriate time before installing and switching ON the sterilizer. Sterilizers arriving from cold locations could contain moisture, affecting the electrical parts and it could lead to unsafe operation for the user if switched ON immediately. The sterilizer must be removed from the box and transported by two people. Lisa 517 40 kg Total weight: Lisa 522 50 kg NOTE:
2.2
Check the external condition of the box and the sterilizer. In case of any damages, immediately contact your dealer or the shipping agent that has carried out the transport.
STANDARD ACCESSORIES
Following accessories are supplied together with the sterilizer: ACCESSORIES PLACED INSIDE THE STERILIZATION CHAMBER Accessory
Description
Tray
Perforated anodized aluminum tray
Reversible rack
Stainless steel rack to accommodate 5 trays or 3 cassettes. Optional racks available (see Appendix 8). For more information contact your dealer.
Quantity 5
1
Tray holder
Tray holder for removing trays from the sterilization chamber.
1
Drain tube
Transparent drain tube with quick connector
1
Mains cable
Length = 2 meters
1
Funnel
Funnel to fill the clean water tank.
1
Wall spacer
Spacer to keep distance between wall and steril.
2
DOCUMENTS PLACED INSIDE THE PACKAGE Declaration of conformity CE
Sterilization chamber:
1
Sterilizer:
1
User manual
Current user manual
1
Warranty card
Warranty card
1
Works Test Report
Norm EN 13060: small steam sterilizers.
1
ACCESSORIES PLACED INSIDE THE PACKAGE Memory card USB reader
External memory card reader
1
Memory card
Reads and saves cycle data
1
3
3.
UNIT DESCRIPTION
3.1
FRONT VIEW
Door seal
Water tank cover
Sliding cover for manual filling of the clean water tank
Chamber door
Sterilization chamber
Touch- screen
Door pin Service door Dust filter
3.2
SERVICE DOOR
Bacteriological filter
Main switch Quick connector for manual draining of used water tank Quick connector for manual draining of clean water tank
4
Connection to unlock the chamber door (service technicians only) Connection for the Air Detector Test Device (optional) Memory card slot
3.3
REAR VIEW
Parallel port (printer) Circuit breaker (manual reset)
Male serial port
Female serial port
Mains plug socket
Safety valve cover Mains cable guide
Test connection
Condenser vent Plug for an external water supply system Air filter
Fitting for an external water supply system
D1
D3
D2
D4 Fitting for continuous draining of used water
5
3.4
DESCRIPTION OF THE INTERNAL WATER TANKS
The sterilizer is equipped with two independent water tanks, one for clean water and one for used water (capacity of 3.5 litres each). The tank on the right side is called "clean water tank" and has to be filled with distilled or demineralized water required for the sterilization process. It is fitted with a minimum (0.6l) and a maximum (3.5l) water level sensor. The tank can be filled manually through the tank hole on the top of the sterilizer or automatically with an external water supply system (Dem 32 or Osmo) connected to point D2 at the back of the sterilizer (see § 4.5.1). Use the quick connector behind the service door (left/blue) to drain the clean water tank for cleaning purposes (see chapter on maintenance). The tank on the left side is called "used water tank" and contains the used water collected at the end of each sterilization cycle. It is fitted with a maximum water level sensor (3.5l). Use the quick connector behind the service door (right/grey) to drain the used water tank (see chapter on maintenance). The used water tank can also be drained automatically by connecting a drain tube to the D4 fitting located at the back of the sterilizer (see § 4.6). Vent Clean water tank Used water tank Water level sensors Water inlet for manual filling of clean water tank
Quick connector for manual draining of clean water tank
Fittings for automatic filling/draining of tanks
Quick connector for manual draining of used water tank
NOTE:
6
The water consumption per sterilization cycle varies depending on the type and weight of the sterilization load. The capacity of the clean water tank is sufficient to run 8 - 12 sterilization cycles.
3.5
CHAMBER RACK Insert the rack into the sterilization chamber, align it at the center/bottom of the chamber and push it gently into position until it clicks. The chamber rack is reversible and can be used for: 5 trays horizontally or 3 cassettes vertically;
or, if inserted in a 90° degree rotated position,
3 trays or 3 cassettes horizontally.
3.6
USABLE SPACE IN THE CHAMBER The chamber usable space is the maximum volume of the chamber for accommodating a sterilization load. This volume is equivalent to a pipe with the following dimensions: Lisa 517 195 x 195 x 297mm (WxHxD); equal to the volume of 11.5 litres
H
W A
D Ritor no alla sch er mata pre ceden te
Lisa 522 195 x 195 x 390mm (WxHxD); equal to the volume of 15 litres The capacity/volume is identical for all sterilization cycles and types of load.
7
4.
INSTALLATION
4.1
SETUP
The sterilizer has been calibrated and intensively tested in the factory prior to shipping. It does not require any calibration during installation. Observe the following environmental conditions: Working temperature range: from +5°C to +40°C / relative humidity: 0 - 90% Storage temperature range: from -20°C to +60°C / relative humidity: 0 - 90% (empty tanks) Install the sterilizer as outlined below: Place the sterilizer on a flat and level surface. Do not place the sterilizer so that it is difficult to open the service door and operate on the controls in it (main switch, microbiological filter, drains) The maximum weight of the sterilizer with a full clean water tank and the chamber fully loaded is: 52 kg (Lisa 517) 63 kg (Lisa 522) Leave a gap of 50mm in the back and 10mm on each side of the sterilizer to ensure adequate ventilation. Mount the supplied wall spacers at the back of the sterilizer (see image to the right). Do not place the sterilizer near a sink or in a location where it is likely to be splashed with water - danger of electric short circuit! Install the sterilizer in a well-ventilated room. Keep the sterilizer away from all sources of heat.
4.1.1
Securing the sterilizer with a safety bracket If the bench on which the sterilizer is installed is small, there is a risk of the sterilizer tilting when the chamber door is opened and leaned upon. Use the safety bracket (optional; see Appendix 9) to avoid tilting.
- Move the two front feet in the rear position. - Drill two holes on the installation surface and fix the bracket with the supplied screws. Wall Screws Ø4,8x22
113 Minimum distance (millimeters)
40
2 holes Ø3x20
Refer to safety bracket installation procedures for further details. Position the sterilizer on the bench with one of the rear feet inserted in the safety bracket.
8
4.2
ELECTRICAL POWER SUPPLY
The electrical power supply of the sterilizer must comply with all applicable standards in the country of use. The following characteristics are required: - Single - phase 200 - 240 VAC ±10%, 50/60 Hz, 10 A, on a dedicated circuit. - Installation category / mains overload category = II - 10 A differential circuit breaker with a sensitivity of 30 mA. The circuit breaker must be a certified type according to applicable norms. - Maximum power consumption of the sterilizer is 2,000 2,400W (10 A) - A grounded connection is essential. Check that the voltage specified on the name plate located on the backside of the sterilizer corresponds to the supplied mains voltage. The overall electrical safety of the sterilizer is guaranteed only if the mains voltage supply is properly grounded according to all applicable norms. If unclear, have the electrical installation checked by a qualified electrician. Do not plug other equipment into the same socket/circuit. Do not bend or twist the mains cable. Only use the original mains cable as supplied with the sterilizer. Do not use cable extensions.
4.3
PRINTER (optional) We recommend the use of LisaPrint (conforms to the IEC 61010-1 norm) as it has been tested for compatibility with the sterilizer and its software. The use of printers other than the ones listed in the relevant printer setup menu (see 6.1.3) clears the manufacturer of any responsibility for warranty or any other claims.
- Connect the printer cable to the 25-pin parallel port socket at the back of the sterilizer. Cable length should not exceed 2 meters.
- Connect the printer mains cable. - Switch ON the printer. - Select the printer type (§ 6.1.3). LisaPrint is the default setting (Print-p). All necessary data to document sterilization cycles is automatically printed (for details see Appendix 8). NOTE:
Lisa 517/522 sterilizers offer the option to digitally save cycle data on removable memory cards.
9
4.4
LISAWARE - CONNECTING THE STERILIZER TO A COMPUTER (optional)
Connect up to four Lisa 517/522 sterilizers to a computer or computer network with LisaWare, a computer software that allows automatic saving of cycle data on a computer and the remote monitoring of the cycle progress on the computer screen. Sterilizers can be connected to a computer in two ways, either via a LAN connection using a MOXA ethernet to serial adaptor, or via a serial connection.
4.4.1
LAN connection W&H recommend using the MOXA NPort 5110 which has been tested and is compatible with the Lisa sterilizer and its software. The use of other ethernet to serial adaptors removes all responsibility from the manufacturer in regards to the correct functioning of the system, the warranty and any other claims. -
Your sterilizer features two serial ports in the back: male and female; either port can be used for data transfer. Switch on the sterilizer, access the setup menu, enter the “RSDATA port selection” option and select one of the two available serial ports for data transfer (see 6.1.17). - Scroll to the “Connection to PC/Log” option in the setup menu and select and confirm “yes” (see 6.1.16). - Please refer to the MOXA configuration sheet for instructions on how to configure your MOXA NPort 5110 to your network. - Once you have configured your MOXA, use one of the supplied serial cables (F-F and F-M) to connect the MOXA to the serial port you selected for data transfer. - Connect the MOXA to the LAN using an ethernet cable (or if connecting directly to the computer use a cross-wire cable). - Use the supplied cable to connect the MOXA to the mains power supply. For further information, please refer to the “Connection Lisa to LAN or PC” quick-start guide. W&H does not accept any responsibility should the MOXA interfere with the normal functioning of your network.
4.4.2
Serial connection -
Your sterilizer features two serial ports in the back: male and female; either port can be used for data transfer. Switch on the sterilizer, access the setup menu, enter the “RSDATA port selection” option and select one of the two available serial ports for data transfer (see 6.1.17). - Scroll to the “Connection to PC/Log” option in the setup menu and select and confirm “yes” (see 6.1.16). - Use one of the supplied serial cables (F-F and F-M) to connect the sterilizer (from the selected serial port) directly to the serial port on your computer. For further information, please refer to the “Connection Lisa to LAN or PC” quick-start guide.
10
4.5
DEMINERALIZER (optional)
W&H Sterilization offers external water supply systems for automated supply of demineralized water to the sterilizer(s) (Dem32 or Osmo). Once installed, the clean water tank no longer needs to be filled manually but will be filled automatically with demineralized water. The Dem32/Osmo features
WATER SUPPLY
resin cartridges to remove minerals from city water to assure a constant supply of good quality water for the sterilization process.
4.5.1 Connecting an external water supply system (demineralizer)
D2 = filling
To install a Dem32/Osmo water supply system, an interface cable and the fittings D2 and D4 located at the back of the sterilizer are used. For more information, refer to the Dem32/Osmo user
Interface cable
manuals.
4.6
CONTINUOUS DRAINING (optional)
4.6.1
Connecting the drain tube DRAIN
The sterilizer can be connected to a drain (or simply to a sink) for continuous draining of the used water tank.
- Unscrew the plug from the D4 fitting at the back of the sterilizer.
- Install the 1/8” barb fitting and the drain tube from the kit and route the tube to a drain or sink (kit ordering number: G0053060). If unclear, have the installation checked by a qualified plumber.
D4 = draining
The drain tube must not be longer than 5 meters. The draining point must be at least 20cm below the surface on which the sterilizer is placed. For further information, refer to the Dem32/Osmo user manuals.
11
5.
GETTING STARTED
5.1
THE USER INTERFACE To power the sterilizer press the main switch located behind the service door (see image to the left). When the sterilizer is switched ON, the chamber automatically heats to 120°C. This process (pre-heating) takes approximately 10 minutes. If a sterilization cycle is initiated during the pre-heating phase, it will start automatically once the pre-heating temperature is achieved. Once the sterilizer is switched on, the following screen appears: Sleep mode
Menu option
Available cycles
Current date
Current time
Selection confirmation
Cycle selection
Cycle counter
To run a sterilization cycle, place the load in the sterilization chamber and close the chamber door. Select a cycle by pressing the “Cycle selection” icon and confirm the selection by pressing the “Selection confirmation” icon.
Start cycle
A screen will inform you about the maximum load weight limits for the cycle you selected. Initiate the cycle by pressing the “Start cycle” icon.
12
5.2
INITIAL WARNINGS AND SLEEP MODE
This message will appear if the sterilizer is switched on and the clean water tank is empty. Fill the clean water tank with distilled or demineralized water (see § 5.4).
NOTE:
If this message is displayed it is not possible to initiate a sterilization cycle. The message disappears automatically once the clean water tank is filled.
If the sterilizer is not used for a certain programmable time period, it will automatically switch to “Sleep mode” to conserve energy (programmable time-out; default 1 hour; see § 6.1.2). Press the “Sleep mode” icon to get back to the main menu. Sleep mode
5.3
DATE-CLOCK SETTING
To properly save or print cycle data on a memory card or on a printout, the internal sterilizer time and date has to be set since these parameters are included in the sterilization cycle data report. Please refer to § 6.1.9 for details on how to properly set the sterilizer date and time.
13
5.4
FILLING THE CLEAN WATER TANK
5.4.1
Manual filling Use only high quality distilled or demineralized water (see Appendix 7).
Slide the tank cover to the right to access the clean water tank inlet. Remove the cap (1) from the tank inlet. Insert the funnel and fill the clean water tank with app. 3.5 litres of distilled or demineralized water. Once the clean water tank is almost full, an audible tone will sound; stop filling. Place the cap (1) to close the tank. Slide the tank cover back into its original position. Drain the used water tank (see § 5.5). The water level in the clean water tank decreases during every sterilization cycle, while the water level in the used water tank is rising. The sterilizer is equipped with an internal water conductivity sensor that constantly monitors the water quality in regards to mineral content. Should the mineral content go beyond a maximum threshold (conductivity >15µS/cm), a warning message is displayed on the screen.
Every time you fill the clean water tank, drain the used water tank (see § 5.5).
5.4.2
Automated water supply (optional) If an automated water supply system is connected to the sterilizer (see § 4.5), the clean water tank gets automatically filled once the water level drops below the minimum level. An internal water conductivity sensor constantly monitors the water quality in regards to mineral content. Should the mineral content go beyond a maximum threshold (>15µS/cm conductivity as defined by the European Norm EN 13060), a warning message is displayed on the screen (see screenshot to the left).
Water of low quality with high mineral contents can impair the sterilization process and seriously damage the internal components of the sterilizer. Damages caused by water of poor quality will not be covered by the manufacturer’s warranty. Once the non conform water message is displayed, check the water source (refer to the water supply system user manual; replace resin cartridges etc.). In case distilled or demineralized water is purchased, switch to a brand of better quality.
14
5.5
DRAINING THE USED WATER TANK When the waste water in the used water tank reaches the maximum level, the following message is displayed:
The message will disappear once the used water tank got drained. Keep draining the tank until it is completely empty.
5.5.1
Manual draining Open the service door at the front of the sterilizer. Insert the drain tube into the quick connector for the used water (grey connector / right). Drain the used water tank until it is completely empty. Press the push-button on top of the quick connector to dislodge the drain tube.
Never reuse water from the used water tank!
5.5.2
Continuous draining
If the permanent drain tube for automated draining of the used water is mounted, the used water tank gets drained automatically. For more information see § 4.6. If the sterilizer is not used for more than 3 days, both water tanks should be completely drained in order to avoid algae growth or any other deposits.
5.6
MEMORY CARD
5.6.1
Inserting / removing the memory card Lisa 517/522 sterilizers are equipped with a digital cycle data recording system. Cycle data are written and saved on removable/rewritable memory cards.
- Insert the memory card into the dedicated slot behind the service door until it clicks into its final position. Ensure that the flat corner of the card points to the top/right (see image to the right).
- Periodically remove the memory card to download cycle data to a computer.
- To remove the memory card, slighty push it in and pull it out gently. For further instructions on the use of the memory card, see §12.
15
6.
PROGRAMMING
Lisa 517/522 sterilizers allow the user to program a number of different features. The following section shows step by step how each of these features can be programmed. Press the menu icon to view the list of available options.
The available menu options are shown below:
Menu
6.1 SETUP MENU The Setup menu permits to set the following options: § 6.1.1 – Language § 6.1.2 – Sleep mode § 6.1.3 – Printer § 6.1.4 – Label printer * § 6.1.5 – Automatic printing * Exit and return to higher menu level § 6.1.6 – Manual printing * § 6.1.7 – Storage time * § 6.1.8 – User name Confirm and return to previous § 6.1.9 – Date-Clock setting menu § 6.1.10 – Date format § 6.1.11 – Clock format § 6.1.12 – Display contrast § 6.1.13 – Display backlight § 6.1.14 – Acoustic tones § 6.1.15 – Night cycle shortcut § 6.1.16 – Connection to PC/Log § 6.1.17 – RSDATA port selection
Scroll up Current selection
Scroll-down
* The sub-menus related to label printing are only available when the label printer LisaSafe is connected.
6.1.1
Language
Scroll up
Use this menu to select the user interface lanExit and return to higher guage. menu level Confirm and return to previous menu
16
Current selection
Scroll down
6.1.2
Sleep mode
In “Sleep mode” the sterilizer reduces the power consumption to a minimum. The sterilizer will switch to sleep mode whenever it is idling without being used. Use this menu to program the time before the sterilizer automatically switches to sleep mode. The time range is from 0 (never in sleep mode) to 8 hours. The factory setting is 1 hour.The time can be set in fixed increments of 10 minutes.
6.1.3
Printer
Use this menu if you want to connect a printer for cycle data recording. Select the printer by using the scroll icons, press “Confirm” to save the configuration. For more information on connecting a printer to your sterilizer, see § 4.3.
6.1.4
Exit and return to higher menu level
Confirm and return to previous menu
Exit and return to higher menu level
Increase time
Decrease time
Scroll up
Current selection Confirm and return to previous menu Scroll down
Label printer (available only when a LisaSafe label printer is connected)
6.1.5 Automatic printing (available only when a LisaSafe label printer is connected) 6.1.6
Manual printing (available only when a LisaSafe label printer is connected)
6.1.7 Storage time/weeks (available only when a LisaSafe label printer is connected) 6.1.8
User name
Use this menu to program the user or the office name. The name will be included in the cycle data report. Select a character by pressing the ”Increment” or “Decrement” icons. Press the “Move next” icon to move to the next position. When the setting is completed, press the “Confirm” icon.
Exit and return to higher menu level User or office name Confirm and return to previous menu
Increment
Move next
Decrement Move previous
6.1.9
Date-Clock setting
Use this menu to set the internal time and date of the sterilizer. It is important to set these parameters as they are included in the cycle data report. Select a character by pressing the ”Increment” or “Decrement” icons. Press the “Move next” icon to go to the next position. When the setting is completed, press the “Confirm” icon.
Exit and return to higher menu level Date
Increment Time Move next
Confirm and return to previous menu
Decrement
17
6.1.10
Date format
Use this menu to change the date format. Press the “Scroll up” and “Scroll down” icons to select the format. Press the “Confirm” icon to save the selection.
Exit and return to higher menu level
Scroll up
Confirm and return to previous screen Scroll down
6.1.11
Clock format
Use this menu to change the clock format. Press the “Scroll up” and “Scroll down” icons to select the format. Press the “Confirm” icon to save the selection.
Exit and return to higher menu level
Scroll up
Confirm and return to previous screen Scroll down
6.1.12
Display contrast
Use this menu to change the screen contrast for a better visualization in regards to the surrounding environment. Press the “Increase contrast” and “Decrease contrast” icons to adjust the contrast. Press the “Confirm” icon to save the selection.
6.1.13
Display backlight
Use this menu to adjust the screen backlight. Press the “Increment” and “Decrement” icons to change the setting. Press the “Confirm” icon to save the selection.
Exit and return to higher menu level
Increase
Confirm and return to previous screen Decrease
Exit and return to higher menu level
Increase
Confirm and return to previous screen
Decrease
6.1.14
Acoustic tones
Use this menu to set the amount of audible tones the sterilizer is generating for various procedures: “All”: All actions come with audible tones. “Reduced”: Most actions come with audible tones. “Minimum”: Only critical actions come with audible tones (e.g., alarms, end of cycle, etc.). Press the “Scroll up” and “Scroll down” icons to select the setting. Press the “Confirm“ icon to save the setting.
18
Exit and return to higher menu level
Scroll up
Confirm and return to previous screen Scroll down