Air Polisher USE AND MAINTENANCE MANUAL
PF-60, PF-60 M, PF-60 N, PF-60 S
Copyright © Mectron S.p.A. 2022. All rights reserved. No part of this document may be reproduced, in any form whatsoever, without the prior written consent of the copyright holder. Trademarks All trademarkes belong to their respective owners: • • • • •
MULTIFLEX® - MULTIFLEX® is a registered trademark of Kaltenbach & Voigt (KaVo®) GmbH company. SIRONA® is a registered trademark of SIRONA DENTAL SYSTEMS GMBH company.
ROTO QUICK® - ROTO QUICK® is a registered trademark of W&H® Deutschland GmbH company. MACHLITE® - MACHLITE® is a registered trademark of Nakanishi Corporation and Nakanishi Shika Kikai (NSK®) Ltd company. PHATELUS® - PHATELUS® is a registered trademark of Nakanishi Corporation and Nakanishi Shika Kikai (NSK®) Ltd company.
I
proxeoAURA CONTENTS
1
Introduction
2
Identification Data
6
3
Delivery
7
4
Installation
10
5
Use
12
6
Cleaning, Disinfection and Sterilisation
21
1.1 1.2 1.2.1 1.2.2 1.2.3 1.2.4 1.3 1.4 1.5 2.1 2.2 3.1 4.1 4.2 5.1 5.2 5.3 6.1 6.2 6.2.1 6.2.2 6.3 6.3.1 6.4 6.5 6.5.1 6.5.1.1 6.5.1.2 6.5.1.3 6.5.1.4 6.5.2 6.5.2.1 6.5.2.2 6.6 6.6.1 6.6.2 6.7 6.7.1 6.7.2 6.8 6.8.1 6.8.2 6.8.2.1
II
Intended Use Description of the Appliance Eligible Patient Group Patient Selection Criteria Indications for Use Users Disclaimer Safety Requirements Symbols Appliance Identification Data Polisher Nozzle Identification Data
List of components Quick Couplings for Turbine Safety Requirements in the Installation Phase Safety Requirements Before and During Use Instructions for Use Cleaning the Powder Container
Preparation Cleaning the Device air polisher Required Materials Cleaning Method Disinfecting the air polisher Device Required Materials Daily Decontamination of the Dental Unit Water Circuit Cleaning the Sterilisable Accessories Manual Cleaning Required Materials Polisher Nozzles K9 Wrench Cleaning Needles Automatic Cleaning Required Materials Cleaning Procedure Cleaning Control Required Materials Procedure for Cleaning Control Drying Required Materials Drying Procedure Sterilisation Preparation Sterilisation Method Special Information
1 1 2 2 2 2 2 3 3 4 6 6
7
11 12 13 14 18
22 23 23 23 24 24 26 26 27 27 28 30 32 34 34 34 35 35 35 36 36 36 36 36 37 37
7
Maintenance
38
8 9 10
Methods and Precautions for Disposal Technical Data Troubleshooting
41 42 43
11
Warranty
47
7.1 7.1.1 7.1.2 7.2 7.2.1 7.2.2 7.3 7.4 7.5
10.1 10.2
Maintenance of the Polisher Nozzle Routine Maintenance of the Polisher Nozzle Clogged Polisher Nozzle Maintenance of the Core Unit Routine Maintenance of the Core Unit Clogged Core Unit Checking the Connector of the Polisher Nozzle Checking the Turbine Quick Coupling Routine Decontamination of the Dental Unit Water Circuit
Quick Troubleshooting Shipping to an Authorised W&H Service Centre
38 38 38 39 39 39 40 40 41
43 46
III
Introduction
1
INTRODUCTION
Carefully read this manual before proceeding with installation operations, use, maintenance or other intervention operations on the appliance. This manual must always be available to the operator. Important: To avoid damages to persons or property, carefully read all the "Safety Requirements" in the manual. Depending on the level of severity, the safety requirements are classified with the following indications: WARNING: (Always referred to damage to persons) CAUTION: (Referring to possible damage to property) The aim of this manual is to make the operators aware of the safety regulations, instal-
1.1
lation procedures, instructions for proper use and maintenance of the appliance and its accessories. Use of this manual for aims other than those strictly linked to the installation, use and maintenance of the appliance is prohibited. The information and illustrations in this manual were updated on the edition date shown on the last page. MECTRON is engaged in continuously updating its products with possible changes to the components of the apparatus. In case you encounter discrepancies between the descriptions found in this manual and the equipment in your possession you can: • • •
check for any available updates in the MANUALS section of the W&H website; ask clarifications to Your Dealer;
contact the W&H After Sales Service.
Intended Use
This device is a water jet, air, and powder polisher for dental use, designed for a complete supraand subgingival prophylaxis treatment, with a specific intended use depending on the type of powder distributed by W&H. Air-polisher for supragingival use: Air-polisher for subgingival use: • • • •
Removal of supragingival bacterial plaque;
Removal of stains on the surface of teeth;
• •
Removal of subgingival bacterial plaque; Removal of Biofilm in the prevention of Peri-implantitis.
Preparation of cavity for better adhesion between enamel and filling material; Prophylaxis on patients undergoing orthodontic therapy.
This air-polisher can be used on patients of any age and gender requiring a tooth polishing treatment. There are no contraindications for specific population segments.
1
proxeoAURA 1.2
Description of the Appliance
This air polisher is a compact water jet, air, and powder dental device used in the dental sector for prophylaxis and tooth polishing treatments. This air polisher can be used with powders of various particle sizes and guarantees a constant, efficient jet of powder and simple, reliable use. The operating principle of a air polisher is based on the mechanical action obtained by a mechanical jet of powders accelerated by a flow of compressed air. The kinetic energy thus impressed on the
1.2.1
particles is dissipated when it hits the enamel surface, thus producing a delicate but efficient polishing action. The action is completed by a jet of water which, by means of the vacuum created around the nozzle, forms a bell-shape around the main flow, thus producing a double effect: to prevent much of the rebound and the leakage of the powder jet and continuously wash the treated area, dissolving the powder itself. This air polisher can be supplied with different quick couplings for connection to the turbine of the dental unit.
Eligible Patient Group
This medical device is designed to be used with the following patient groups: •
Children;
•
Adults;
• •
Adolescents;
This medical device can be used on patients of any age, weight, height, gender, and nationality, where applicable.
Seniors.
1.2.2
Patient Selection Criteria
It is not recommended to use the device in the following cases: 1. 2. 3.
Upper respiratory tract infections, chronic bronchitis/asthma;
Pregnant and breastfeeding women; Patients undergoing treatment
4.
(radiotherapy, chemotherapy, antibiotics); Acute infectious oral lesions.
It is not recommended to use the powders in the following cases: 1. 2.
Allergy to the aroma of the powder;
Patients on a sodium restricted diet or who suffer from serious respiratory
problems, such as chronic bronchitis, asthma, emphysema, etc., unless otherwise specified by their doctor.
All air-polishing devices are intended for professional use only. The user is therefore the only person able to decide if and how to treat their patients.
1.2.3
Indications for Use
1.2.4
Users
The device is suitable for use on all eligible patients (see Chapter 1.2.1 on page 2) whose doctor has prescribed a professional tooth polishing treatment, as part of the intended use of the device (see Chapter 1.1 on page 1). The device must be used exclusively by specialised and properly trained personnel, specifically the dental surgeon/dentist or dental hygienist, who must be able-bodied, adult, of any weight, age, height, gender, and nationality.
2
Introduction 1.3
Disclaimer
The manufacturer Mectron disclaims all liability, express or implied, and cannot be held responsible for direct or indirect personal injury and/or property damage, occurring as a result of incorrect procedures linked to the use of the appliance and its accessories. The manufacturer Mectron cannot be held liable, expressly or implicitly, for any type of personal injury and/or property damage, inflicted by the user of the product and its accessories and which occurs in the following cases: • •
• •
Misuse or use during procedures other than those specified in the destination of use of the product; The environmental conditions for preservation and storage of the device are not complying with the requirements indicated in Chapter 8 on page 41; The appliance is not used in accordance with all the instructions and requirements described in this manual;
Assembly operations, extensions, readjustments, upgrades and repairs of the device are carried out by personnel not
1.4
•
• • • • •
authorised by Mectron or by W&H;
Misuse, abuse, abnormal use, negligent use, intentional misconduct or use exceeding the limits of the appliance indicated and allowed and/or normal wear or deterioration, ill-treatment and/ or incorrect interventions; Any attempt to tamper with or modification of the appliance under every circumstance;
Shortage of stock material (handpiece, nozzles, wrenches) to be used in case of a standstill due to faults or problems;
Incorrect/omitted maintenance compared to what is stated in Chapter 7 on page 38 of this manual; Breach of the requirements and the information contained in Chapter 6 on page 21 of this manual;
Unauthorised repairs in accordance with the indications contained in Chapter 10.2 on page 46 this manual.
Safety Requirements
CAUTION: It is prohibited to make any changes to this appliance. WARNING: The appliance must be used in a dental practice or clinic, or in professional oral hygiene and preventive care centres. WARNING: Qualified and specialist personnel. The appliance should only be used by qualified personnel with adequate medical culture; for the use of the device are not requested special training activities. Use of the device does not produce any side effects if used correctly. Misuse may cause soft tissue damage and tissue emphysema.
WARNING: Intended use. Use the device only for the destination of use for which it is provided (see Chapter 1.1 on page 1). Failure to observe this regulation can cause serious injury to the patient, the operator and damage/faults on the device. WARNING: Contraindications. Patients wearing contact lenses must remove them prior to receiving treatment with the jet polisher. WARNING: Contraindications. Do not direct the air/powder/water jet towards the eyes. Failure to comply with this indication may cause serious eye damage. It is advisable for all involved persons to wear safety goggles (patient, hygienist, and dentist).
3
proxeoAURA WARNING: Contraindications. Do not aim the air/powder/water jet towards composites, fillings, and bridges. Failure to comply with this indication may case damage thereto. WARNING: Contraindications. Do not aim the air, specific supragingival powder, or water jet onto the soft tissues or inside the gingival sulcus. Failure to comply with this prescription can cause a gingival tissue emphysema (emphysema of the mucous and/or subcutaneous). WARNING: Contraindications. Do not use the device near areas subject to recent dental extraction and in traumatised/ damaged areas (or areas nearby) due to the risk of emphysema. WARNING: Contraindications. It is advisable for the dentist and hygienist to use a protective face mask to limit the inhalation of powder. WARNING: Checking the condition of the device before treatment. Before every treatment always check the perfect operation of the appliance and efficiency of the accessories. If operation anomalies are observed do not carry out the treatment. If the faults concern the appliance, contact an Authorised W&H Service Centre.
WARNING: Cleaning, disinfection, sterilisation of new or repaired products. All new or repaired products are delivered non-sterile. Before the treatment, any new or repaired products must be cleaned, disinfected and sterilised, strictly following the instructions provided in Chapter 6 on page 21. WARNING: Infection control. For utmost patient and operator safety, use only clean, disinfected and sterilised accessories. Strictly follow the instructions provided in Chapter 6 on page 21.
CAUTION: Contraindications. After having sterilised the polisher nozzles and any other sterilisable accessory, wait for them to completely cool down before reusing them. WARNING: Use only original W&H accessories and spare parts.
WARNING: In case of knocks, damage, or dents on the clear air polisher core unit, send it to an Authorised W&H Service Centre for analysis and inspection. WARNING: In case of an adverse event and/or accident attributable to the device during correct use and in accordance with the intended use, a report must be made to the Competent Authority and manufacturer indicated on the product label. NOTE: To check and guarantee the perfect operation of the device and prolong its service life, maintenance should be carried out by an Authorised W&H Service Centre on an annual basis.
1.5
Symbols Symbol
0051
Description Device compliant with Regulation (EU) 2017/745. Notified body: IMQ S.p.A. Caution! Read the instructions for use
4
Symbol
Description Medical device
Operating instructions
Introduction Symbol
QTY.1
Description
Symbol
Description
Manufacturer
Date of manufacture
Serial number
Batch number
Product code
Disposable
Expiry date
Sterilised with Ethylene Oxide (EO)
Not sterile
The sterilisable materials must be autoclaved and can resist up to a maximum temperature of 135 °C
Treatable in thermal-disinfector
The appliance and its accessories must not be disposed of or treated as municipal solid waste
Generic warning sign a)
Temperature limits for transport and storage
Moisture limits for transport and storage
Atmospheric pressure limits for transport and storage
Quantity in package: 1
Do not use if the package is damaged
For US market only Rx only
Table 1 – Symbols
CAUTION: US federal law restricts sale to orders by licensed dentists or dental hygienists only.
a) The symbol is represented by a yellow triangle and black graphic symbol.
5
proxeoAURA
2
IDENTIFICATION DATA
A correct description of the model and of the serial number of the appliance will allow the After Sales Service to provide fast and effec-
2.1
tive answers. Always provide this information every time you contact a W&H Service Centre.
Appliance Identification Data
The appliance name, serial number and other information (Figure 1) is laser engraved on each device. The complete technical specifications are provided in Chapter 9 on page 42.
Data Matrix HIBC 128
SN 2 1 6 0 0 0 00
SN 2 1 60 0 00 0
0051
perio
PF-60
0400011X YYYY-MM
made in italy
0051
perio
prophy
PF-60
0400011X YYYY-MM
made in italy
Figure 1 – Appliance Identification Data
Polisher Nozzle Identification Data
0051
1
120°
2
3 00000000 00000000
The name of the polisher nozzle itself (Figure 2 - Ref. 1), W&H logo (Figure 2 - Ref. 2), the batch number of the polisher nozzle (Figure 2 - Ref. 3), the HIBC data matrix (Figure 2 - Ref. 4) and other symbols described in Chapter 1.5 on page 4 are laser engraved on each polisher nozzle.
120°
2.2
4
YYYY-MM
Figure 2 – Polisher
6
Nozzle Data
Identification
Delivery
3
DELIVERY
3.1
List of Components
The air polisher is supplied in two base sets: •
Prophy-Set;
•
Paro-Set.
Prophy-Set Supply Item
Device core unitc)
Code
Description
04000123
W&H® coupling / PF-60 (ROTO QUICK®)
04000124
KaVo® coupling / PF-60 M (MULTIflex® - MULTIflex® LUX)
04000125
NSK® coupling / PF-60 N (PHATELUS® - MACHLITE®)
04000126
Sirona® coupling / PF-60 S (SIRONA® R/F - SIRONA® B)
Ref.b)
A
Cleaning needle Ø 0.4 mm
02900050
B
Cleaning needle Ø 0.8 mm
02900121
C
Pipe cleaner
02900131
D
Use and Maintenance Manual
02150396
Information: Online Documentation
02150660
O-ring replacement KIT
02900141001
F
Cap and cap O-ring replacement KIT
02900142001
G
Jar of specific supragingival powder
03140030
Prophylaxis Powder: SOFT + (single package)e)
L
TERMINALE PULITORE LUNGO 120°
03340017
120° long air-polishing nozzle
H5
E
b) See image on front cover. c) The air polisher device can be connected to the dental unit turbine using different couplings (see Chapter 4.1 on page 11). A different type of device core unit will be supplied depending on the requested coupling.
7
proxeoAURA Paro-Set Supply Item
Device core unitc)
Code
Description
04000123
W&H® coupling / PF-60 (ROTO QUICK®)
04000124
KaVo® coupling / PF-60 M (MULTIflex® - MULTIflex® LUX)
04000125
NSK® coupling / PF-60 N (PHATELUS® - MACHLITE®)
04000126
Sirona® coupling / PF-60 S (SIRONA® R/F - SIRONA® B)
Ref.b)
A
Cleaning needle Ø 0.4 mm
02900050
B
Cleaning needle Ø 0.8 mm
02900121
C
Pipe cleaner
02900131
D
Use and Maintenance Manual
02150396
Information: Online Documentation
02150660
O-ring replacement KIT
02900141001
F
Cap and cap O-ring replacement KIT
02900142001
G
Jar of specific supragingival powder
03140030
Prophylaxis Powder: SOFT + (single package)e)
L
Jar of specific subgingival powder
03140024
Prophylaxis Powder: SENSITIVE + (single package)e)
K
TERMINALE PULITORE LUNGO 03340017 120°
120° long air-polishing nozzle
H5
TERMINALE PULITORE LUNGO 03340018 PERIO
PERIO air-polishing nozzle
H4
40 pc set of SUBGINGIVAL PERIO TIPS
I
SUBGINGIVAL PERIO TIPS
02900139001
K9 Wrench
02900140001
8
E
J
Delivery Item
Accessories
Code
TERMINALE PULITORE CORTO 03340021 PERIO
Description
PERIO short air-polishing nozzle
Ref.b)
d)
H1
TERMINALE PULITORE CORTO 03340020 120°
120° short air-polishing nozzle d)
H2
TERMINALE PULITORE CORTO 03340019 90°
90° short air-polishing nozzle d)
H3
TERMINALE PULITORE LUNGO 03340018 PERIO
PERIO air-polishing nozzle d)
H4
TERMINALE PULITORE LUNGO 03340017 120°
120° long air-polishing nozzle d)
H5
TERMINALE PULITORE LUNGO 03340016 90°
90° long air-polishing nozzle d)
H6
SUBGINGIVAL PERIO TIPS
02900139001
40 pc set of SUBGINGIVAL PERIO TIPS d)
I
K9 Wrench
02900140001
d)
J
03140024
Prophylaxis Powder: SENSITIVE + (single package) e)
K
03140023
Prophylaxis Powder: SENSITIVE + (2x160g package) e)
03140030
Prophylaxis Powder: SOFT + (single package) e)
03140029
Prophylaxis Powder: SOFT + (4x250g package) e)
03140016
Prophylaxis Powder: SMOOTH (single package) d)
03140015001
Prophylaxis Powder: SMOOTH (4x250g package) d)
Jar of specific subgingival powder
Jar of specific supragingival powder
L
Includes: Cleaning kit
Cap and cap O-ring KIT d) Produced by Mectron. e) Distributed by Mectron.
02900143001 02900142001
•
Cleaning needle Ø 0.4 mm;
•
Pipe cleaner. d)
•
d)
Cleaning needle Ø 0.8 mm;
M
G
NOTE: Both the items included in the standard supplies and all accessories can be ordered separately by the client.
9
proxeoAURA The device packaging is sensitive to impact, therefore caution must be exercised during its transport and deposit. Do not stack multiple boxes so as not to crush the underlying packages. All material sent by Mectron is checked at the time of dispatch. The device is dispatched with suitable protection and packaging.
4
When the device is received, check the condition of the packaging and in case of damage, deterioration, or signs of tampering, submit a claim to the carrier. Keep the packaging in case any items need to be sent to an Authorised W&H Service Centre and to store the appliance during long periods of non-use.
INSTALLATION
The air polisher device, in order to operate must be connected to a dental unit equipped for water and air distribution.
CAUTION: Check that the dental unit to which air polisher is connected satisfies the specified requirements.
NOTE: The device does not require any electrical and/or electronic connections to be made.
The device must be installed in a suitable place allowing its comfortable use. For the air polisher device to operate correctly, the air and water pressure values must be within the intended range of use, specifically:
The air polisher device is supplied with a specific adapter for the main couplings to the dental unit turbine (see Chapter 4.1 on page 11). CAUTION: Use the air polisher device only with the specific coupling for the relative turbine. Connecting the air polisher device to a connector not intended for use in combination with the dental unit may damage the device. The dental unit to which air polisher is connected must be a medical device that complies with general and applicable product regulations for the specific category of dental unit, and in general, satisfy the following standards: • • •
ISO 7494 Dentistry – Stationary dental units and dental patient chairs; ISO 9168 Dentistry – Hose connectors for air driven dental handpieces;
ISO 60601-1 Medical electrical equipment – Part 1: General requirements for basic safety and essential performance.
• •
For water, from 1.0 to 3.0 bar (from 1000 hPa to 3000 hPa), max 40 °C;
For air, from 2.2 to 3.0 bar (2200 hPa 3000 hPa).
CAUTION: To ensure proper and longlasting performance, ensure the air flowing from the dental unit is thoroughly dry and filtered. Moreover, it is essential to ensure that the air pressure is between 2.2 and 3.0 bar inclusive, measured at the point of the connection between the device and the dental unit. NEVER use excessive air pressure. Increasing the air pressure does not correspond to an increase in performance; to the contrary, the device may become damaged or fail to work properly.
To ensure the correct operation of the device, it is necessary to check: • • •
10
The “spray” water supply coming from the turbine;
The air supply coming from the turbine; The presence of powder.
Installation 4.1
Quick Couplings for Turbine
The quick couplings with which the air polisher device can be supplied are listed in Table 2 on page 11. To connect the device to a quick coupling, fully insert the connector until you hear a “click”. To disconnect the device, follow the instructions in Table 2 on page 11. Quick Coupling
Description W&H® ROTO QUICK®
NSK® MACHLITE® / PHATELUS®
Extraction Procedure On the quick coupling, make the ring slide towards the cord and disconnect the device by rotating it slightly. Do not pull or grasp the cord.
KaVo® MULTIflex®
KaVo® MULTIflex® LUX
SIRONA® B
Firmly grasp the quick coupling and disconnect the device by rotating it slightly. Do not pull or grasp the cord.
SIRONA® R/F Table 2 – Quick Couplings for Turbine
11
proxeoAURA 4.2
Safety Requirements in the Installation Phase
CAUTION: Check that the turbine connector is clean and that its O-rings are in good condition. WARNING: Do not press the turbine control pedal while checking the connector. WARNING: Do not press the turbine control pedal while making the connection with the device. CAUTION: In case of a turbine with light, switch it off.
5
CAUTION: The polisher nozzle and turbine connector must be perfectly dry. Any humidity may cause the obstruction of the air/powder circulation channels. CAUTION: Do not expose the appliance to direct sunlight or sources of UV light. CAUTION: The appliance can be transported but must be handled with care when moved.
USE
Suitable powder and polisher nozzles must be used depending on the type of treatment to be carried out: •
For applications with specific supragingival powders distributed by W&H, it is necessary to:
•
For applications with specific subgingival powders distributed by W&H, it is necessary to:
1. Set the adjustment ring to prophy;
1. Set the adjustment ring to perio;
3. Insert the 120° polisher nozzle on the handpiece.
3. Insert the desired polisher nozzle on the handpiece (120° or PERIO);
2. Fill the powder container with the appropriate powder without exceeding the indicated maximum level;
2. Fill the powder container with the appropriate powder without exceeding the indicated maximum level;
4. In case of subgingival applications and therefore when using the PERIO polisher nozzle, the single-use SUBGINGIVAL PERIO TIP must be inserted.
CAUTION: The SUBGINGIVAL PERIO TIP can only be used with the PERIO polisher nozzle.
CAUTION: Use only and strictly the powders intended for ‘subgingival’ use with the PERIO polisher nozzle.
WARNING: The SUBGINGIVAL PERIO TIPs are single-use sterile. The SUBGINGIVAL PERIO TIPs are supplied sterile and for single use. They must be used on one patient only during a single application, after which they must be removed and disposed of.
CAUTION: Before filling the powder container with the selected powder, make sure it is clean and free of any powder residue from the previous treatment.
12
Use 5.1
Safety Requirements Before and During Use
WARNING: Contraindications. Patients on a sodium restricted diet must not receive treatment with sodium bicarbonate powders. WARNING: Contraindications. Patients who suffer from serious respiratory problems such as chronic bronchitis, asthma, emphysema, etc., must not receive prophylaxis treatment unless otherwise indicated by their doctor. WARNING: Contraindications. Patients wearing contact lenses or glasses must remove them prior to receiving treatment with the air polisher. WARNING: Contraindications. Do not direct the air/powder/water jet towards the eyes. Failure to comply with this indication may cause serious eye damage. It is advisable for all involved persons to wear safety goggles (patient, hygienist, and dentist). WARNING: Contraindications. It is advisable for the dentist and dental hygienist to use a protective face mask to limit the inhalation of powder. WARNING: Contraindications. Do not aim the air/powder/water jet towards composites, fillings, and bridges. Failure to comply with this indication may case damage thereto. WARNING: Contraindications. Do not use the device near areas subject to recent dental extraction and in traumatised/ damaged areas (or areas nearby) due to the risk of emphysema.
WARNING: Checking the condition of the device before treatment. Before every treatment always check the perfect operation of the appliance and efficiency of the accessories. If operation anomalies are observed do not carry out the treatment. If the faults concern the appliance, contact an Authorised W&H Service Centre. WARNING: Use only original W&H accessories and spare parts. WARNING: Infection, water and air circuit control. First use: All parts and reusable accessories (new or returned from an Authorised W&H Service Centre) are delivered in NON-STERILE conditions and must be treated before each use, according to the instructions provided in Chapter 6 on page 21. Subsequent uses: After each treatment, clean, disinfect and sterilise all parts and reusable accessories according to the instructions in Chapter 6 on page 21. CAUTION: Do not attempt to unscrew the cap of the powder container when it is pressurised. CAUTION: Make sure the powder container is perfectly dry. Any humidity may cause the channels to become clogged. WARNING: During the intervention on the patient, do not perform any maintenance tasks on the system.
WARNING: Contraindications. Do not aim the jet of air/powder/water onto the soft tissues or inside the gingival sulcus. Failure to comply with this prescription can cause a gingival tissue emphysema (emphysema of the mucous and/or subcutaneous).
13
proxeoAURA 5.2
Instructions for Use
Select the type of polisher nozzle depending on the treatment to be carried out;
1
2
WARNING: Before performing the subsequent operations, check that the polisher nozzle is properly inserted on air polisher (the ring and tip must be in contact) and that the ring has returned to the initial position.
If the PERIO polisher nozzle has been selected, insert the single-use SUBGINGIVAL PERIO TIP on the front part, pushing it all the way in until it comes into contact. WARNING: Make sure the singleuse SUBGINGIVAL PERIO TIP is properly inserted on the polisher nozzle; the two pieces must be in contact. CAUTION: Use only original W&H accessories. CAUTION: Handle the single-use SUBGINGIVAL PERIO TIP with care. CAUTION: With the PERIO polisher nozzle, only specific subgingival powders distributed by W&H must be used.
14
3
PERIO
Grab the air polisher device, slide the ring down and hold it in this position, insert the selected nozzle until it comes into contact, then release the ring;
120°
CAUTION: The SUBGINGIVAL PERIO TIP can only be used with the PERIO polisher nozzle.