W & H Dentalwerk

Reprocessing Devices

Assistina Weekly Checklist Rev 001

Checklist

1 Page

Weekly checklist Check liquid level! Refill if necessary! Service oil: 500 ml – 02675910  MAX. MIN.  Cleaning solution: 1000 ml – 02680200  Set 02680500: 1000 ml Cleaning solution and 500 ml Service oil  Service oil  Cleaning solution 1  4  Check function!  5  2  3  6  Check o-rings!  O-Ring  02060200  O-Ring (2x)  02060100  O-Ring (2x)  02695700  Look for right position and / or damage!  Adaptor 02690400  Filter  02670500  02 67 05 0  0  After consumption of one 500 ml – bottle of Service oil (approx. 7000 cycles)  W&H Dentalwerk Bürmoos GmbH Ignaz-Glaser-Straße 53, Postfach 1 5111 Bürmoos, Austria  t + 43 6274 6236-0 f + 43 6274 6236-55 office@wh.com wh.com  Changing of filters!  Filter  02675200  Once a year  Form-Nr. 50390 AEN Rev. 001 / 06.04.2020 Subject to alterations  Check connection is tight!