Operating Instruction Manual
26 Pages
Preview
Page 1
Women's Health Stand Alone Colposcope
TM
VideoPath
Colposcope
Operating Instruction Manual
Table of Contents Conventions... 2 General Precautions ... 2 Symbols ... .3 Summary of Warnings and Cautions ... 3 Components ... 4 VideoPath h Colposcope ... .4 VideoPath Light Source ... 4 Optional Accessory ... 4 Nomenclature ... 5 VideoPath Colposcope Front View ... 9 VideoPath Colposcope Side View ... 10 VideoPath Colposcope Back View ... 6 Vertical Pole Stand ... 6 Swing Arm Stand ... 7 .8 Monitor ...
Light Source ... 9 Connection Diagrams ... 11 Connecting VideoPath VLX-10/20 Light Source with Monitor.. ... ..11 Connecting VideoPath VLX- 10/20 Light Source with VCR/Video Printer ... .11 Connecting VideoPath h Colposcope with Light Source.. ... .12 Preparation for Use.. ... .13 General Precautions ... 13 h Colposcope ... .13 Assembly of VideoPath Vertical Pole Stand Assembly ... .13
Swing Arm Stand Assembly... 14 h Colposcope... 15 Setting Up VideoPath Operation... ,... 16 Power Switch... 16 Focus and Zoom Controls...*... 16 Green Filter Controls... 16 Illumination Beam Director... 16 Vertical Height Adjustment Ring... 16 Swing Arm Height Adjustment... 16 Right or Left Positioning of Swing Arm Colposcope... 16 Positioning VideoPath Colposcope... 16 Remote Video Functions... 17 Storage of the VideoPath Colposcope Connector Terminal...17 Maintenance...
.18
VideoPath Colposcope Cleaning... 18 Disinfecting Solutions... 18 Illumination Lens Cleaning... 18 Camera Lens Cleaning... 18 Service Information... 19 Troubleshooting... 20 Specifications... 21
Thank you for purchasing the Welch Allyn VideoPath Colposcope. Follow the operation and maintenance instructions found in this manual and your VideoPath Colposcope will provide you with many years of reliable service. Please read these instructions thoroughly before attempting to use your new VideoPath Colposcope. CAUTION: Federal law restricts sale of this device to, or on the order of, a physician or other appropriately licensed medical professional.
IMPORTANT: The material outlined in this manual should be reviewed and understood prior to operation of the equipment. INDICATIONS FOR USE: For examination of the tissues of the vagina, cervix, and external genitalia, to investigate, by means of magnification, abnormal cervical cytology or suspicious lesions of the lower female genital tract. Also used for corresponding biopsy and treatment, when indicated. NOTE: This device complies with the guidelines of the American Conference of Governmental Industrial Hygienists’ threshold limits values for ultraviolet radiation for exposure time and distance consistent with the intended use of this device.
1
Conventions
CAUTION: Cautions indicate a potentially hazardous situation that, if not avoided, may result in minor or moderate injury and/or damage to the equipment. They also alert against unsafe practices. NOTE: Notes provide supplemental information to the text and indicate a potentially hazardous situation, which, if not avoided, may result in property damage. They also highlight important information on the use of this equipment.
General Precautions . For safety, the VideoPath Colposcope should only be connected to either Welch Allyn VideoPath model number VLX-10 or VLX-20 Light Sources. . The VideoPath Colposcope should not be operated in the presence of flammable or explosive gases or chemicals, or installed in areas where these materials are commonly used. . Keep all liquids away from electrical equipment to avoid the possibility of shock and instrument damage. . Occasionally inspect the umbilical cord and plug for signs of cuts, abrasions or other damage. . The VideoPath Colposcope should never be stored or operated in areas where it could get wet or could be exposed to any environmental conditions such as extreme temperature or humidity, direct sunlight, dust, etc. . The illumination light is extremely bright. DO NOT stare directly into illumination beam director when the Light Source lamp is lit. . All service to the VideoPath Colposcope must be performed by Welch Allyn or by an authorized repair center. . VideoPath Colposcope user should adhere to the operating conditions found in this manual and the applicable Light Source manual. Otherwise, instrument damage may occur and/or operator/patient safety may be compromised. . There are no user serviceable parts in this unit or in its accessories. Any attempt to disassemble and/or repair this unit will result in voiding of the warranty. . Do not immerse any portion of the VideoPath Colposcope.
2
Symbols On: Power: Connects to the low voltage supply.
cl
Off: Power: Disconnects from the low voltage supply.
A
Caution: Consult user’s manual for additional information.
0
6
!
A
A XJ-
Warning: Consult user’s manual for additional information.
Warning: High-intensity light
Summary of Warnings and Cautions
3
Components VideoPath Colposcope The video colposcope includes . . .
VideoPath Colposcope and swing arm stand models 89100/89110 Swing arm hardware kit: Allen wrench Mounting washer Bolt .
VideoPath Colposcope and vertical stand models 88100/88110 Vertical stand hardware kit: Allen wrench Mounting cap
l
l
l
l
l
VLX-10/VLX-20 Light Source Includes: 40500 45500
Light Source (VLX- 10) Light Source (VLX-20)
45510
Water Bottle (Not applicable for VideoPath Colposcope)
761076-0
Power Cord
45512
S-Video Cable
45520
Cleaning/Disinfection Bottle (Not applicable for VideoPath Colposcope)
04060
Lamp Replacement Assembly
65158
RS232 Interface Cable
Optional Accessory
NOTE: Only accessories and components indicated in this manual are to be used with the Welch Allyn Video Colposcope system. 4
Nomenclature VideoPath Colposcope Front View
Name
Function
1. Objective lens
The lens that establishes the magnification and field of view.
2. Illumination beam director
Directs light beam.
VideoPath Colposcope Side View
Name
Function
3. Ventilation slots
Allow air to enter and leave unit, cooling internal components.
4. Zoom control button
Pressing + or - increases or decreases magnification accordingly.
5. Focus control button
Pressing < or > adjusts the focus.
6. Green filter button
Pressing turns the electronic green filter on or off.
7. LED indicator
Illuminates if the electronic green filter is on. Flashes during activation or deactivation.
5
VideoPath Colposcope Back View ~
9-
11
10
Name
Function
8. Power switch
Power control for the VideoPath Colposcope.
9. Ventilation slots
Allow air to enter and leave unit, cooling internal components.
10. VideoPath Colposcope handle
For positioning the colposcope in the proper up/down, left/right position.
11. Remote video function buttons
Allow use of remote video printer functions. (When equipped. Models 88110, 89110 only)
Vertical Pole Stand
Name
Function
12. Vertical rolling base stand
Allows mobility with 5-caster base, including two locking casters.
13. Colposcope connector terminal cord
Couples colposcope camera to colposcope terminal connector.
14. Colposcope connector terminal
Couples colposcope to Welch Allyn Light Source Model VLX- 10 or VLX-20.
15. Colposcope connector terminal hook
Storage hook to support umbilical cord.
16. Colposcope connector terminal hanger Storage hanger to support colposcope connector terminal. 17. Height adjustment ring
Allows vertical movement and locking of stand at desired height; vertical height adjustment 36”-46” (91.4 cm to 116.8 cm).
6
Swing Arm Stand
18
Name
Function
18. Swing arm rolling base stand
Allows overhead positioning of the instrument with 5-caster weighted base, including two locking casters; vertical height 29.5” to 49.5” (74.9 cm to 125.7 cm).
19. Stop pin
Prevents positioning of colposcope in a configuration that limits lateral adjustment.
20. Light Guide
Transmit light from the Light Source lamps to and through the illumination beam director.
21. Electrical Contacts
Electrical connection from the light source to the VideoPath Colposcope.
7 7
Monitor NOTE: The following is a brief overview of the nomenclature of the Sony PVM-1350 Video Monitor. For detailed instructions, please refer to the operation manual that is provided by the manufacturer.
Front Panel of Monitor
Back Panel of Monitor
32
i
30
29 28 26 24
PLEASE REFERENCE YOUR SONY MONITOR MANUAL FOR FURTHER INFORMATION. Name
Function
22. Power
Press to turn the monitor on. The indicator will light up in green.
23. Volume
No function.
24. Contrast
Turn clockwise to increase contrast and counterclockwise to decrease it.
25. Phase (tint)
Turn clockwise to make the skin tones greenish and counterclockwise to make them purplish.
26. Chroma
Turn clockwise to make the color intensity higher and counterclockwise to make it lower.
27. Bright (brightness)
Turn clockwise for more brightness and counterclockwise for less.
28. Aperture (sharpness)
Turn clockwise for more sharpness and counterclockwise for less.
29. Menu button
On-screen programmable controls of the picture.
30. Degauss button
Press this button momentarily. The screen will be demagnetized. Wait for 10 minutes or more before activating this button again.
31. Input select buttons
Press (light on) to select the program to be monitored. A: For signal fed through the LINE A connector. B: For signal fed through the LINE B connector. C: For signal fed through the LINE C connector. (This is the Y/C (S-Video) input signal.) RGB: For signal fed through the RGB connectors.
32. Power supply cord receptacle
Couples with power cord, which should be plugged into a 110-120 VAC hospital grade outlet (220-240 volt, 50 cycle for international).
33. Line A, line B connectors
Two groups (A and B) of line input connectors for composite video.
34. Line C connector Y/C (S-Video)
High resolution S-Video input.
NOTE: This monitor is shipped with default picture settings. Any change of the on-screen controls may result in poor picture quality. 8
Light Source (Front View)
37
35
36
38
41
35. Main power switch
Primary power control for the Light Source. If not activated, other controls will not function.
36. Air pump switch
Not applicable to VideoPath Colposcope.
37. Lamp ignition
Activates the lamps when switch is depressed and main power switch is on.
38. Lamp standby indicator
When LED blinks, lamps cannot be ignited even if lamp ignition switch is depressed.
39. Colposcope connector port
Accepts VideoPath Colposcope umbilical plug to transmit light and electrical signals to the colposcope.
40. Air output port
Not applicable to VideoPath Colposcope.
41. Picture brightness control
Not applicable to the VideoPath Colposcope.
9
9
Light Source (Back View)
48 43 44 45 / /
I
46
42. Power supply cord receptacle
Couples with power cord, which should be plugged into a 115V hospitalgrade outlet.
43. S-Video
Outputs S-Video or Y/C, 4 pin DIN. (Dual connection, VLX 20 only)
44. Composite
Outputs NTSC (composite video), BNC connector. (Not applicable with the VideoPath Colposcope)
45. Earth ground terminal connector
Provides a terminal that is connected to earth ground through the VLX-20 line cord (1/4-36 UNS-2A thread). (VLX 20 only)
46. Equipotentiality terminal connector
Accepts ground connector from electrosurgery equipment to provide a secondary or redundant ground path. (VLX 20 only)
47. RS-232 communication connector port Communication port for use with video printers. (VLX 20 only) 48. Auxilliary port
No function. Future capability. (VLX 20 only)
10
10
Connection Diagrams Connecting VideoPath VLX-10/20 Light Source with Monitor The diagram below illustrates an example of recommended coupling between the VLX-10/20 and a monitor.
S-Video Out
Connecting VideoPath VLX-10/20 Light Source with VCR/Video Printer The diagram below illustrates an example of recommended coupling between the VLX-10/20 and peripheral equipment.
I
, S-Video In
I
Monitor
NOTE: The RS-232 cable is required for the Function Control Buttons to perform properly.
11
Connection Diagrams Connecting the VideoPath Colposcope with Light Source
12
Preparation For Use General Precautions Make sure the unit is always grounded and secure during use. Do not disable power cord ground connection. Grounding reliability is achieved only when power cord is connected to a hospital-grade receptacle. Inspect the electrical plug and cord routinely. Do not use if damaged. Do not use the VideoPath Colposcope in the presence of any flammable anesthetics. Do not open the VideoPath Colposcope housing. An electrical shock hazard exists due to high voltage. There are no user serviceable parts inside the VideoPath Colposcope. NOTE: Opening of the VideoPath Colposcope housing by an unauthorized repair facility will void the product warranty.
Assembly of VideoPath Colposcope Before initial set up of the VideoPath Colposcope, check all components received against the parts list of components (see Components section of this manual) to verify a complete set. If parts are missing, please notify Welch Allyn. Review the Nomenclature, Preparation for Use, Operation and Maintenance sections to become familiar with the equipment.
Vertical Pole Stand Assembly The VideoPath Colposcope stand is shipped unassembled. Minimal assembly is required. 1. To assemble the stand, remove the stand and base parts from their cartons. 2. Place base upside down on the floor and insert casters into holes on the bottom of the base. (Do not place two locking casters next to each other.) Locking caster 49
‘* I
.lP-
=
Locking caster
3. Place the base on the floor with casters down. Lock the two locking casters. 4. Place pole in the base and screw the mounting cap onto the end using the enclosed Allen wrench. Be sure to tighten securely. (The assembly may need to be tilted.)
5. Before using the VideoPath Colposcope, refer to “Setting Up VideoPath Colposcope” section of this manual.
13
Swing Arm Stand Assembly The VideoPath Colposcope stand is shipped unassembled. Minimal assembly is required. NOTE: Base weighs 58 lbs. Assistance may be required to lift. 1. To assemble the stand, remove the stand and base parts from the cartons. 2. Place base upside down on the floor and insert casters into holes on the bottom of the base. (Do not place two locking casters next to each other.) Locking caster +/i/i
I I I
9
Locking caster eiir
a 0 ‘?W?!!!F 3. Place the base on the floor with casters down. Lock the two locking casters. 4. Place the vertical pole section of the assembly into the base. 5. Locate the packaged mounting washer and bolt. 6. Place the bolt through the washer into the bottom of the pole. (The assembly may need to be tilted.)
7. Tighten securely with the enclosed Allen wrench. 8. Remove the bolt from the arm of the colposcope assembly and insert it through the hole near the top of the pole. 9. Position the pole and the arm of the remaining colposcope assembly in close proximity.
10. Tighten the bolt securely with the enclosed Allen wrench.
/
11. Before using the VideoPath Colposcope, refer to “Setting Up VideoPath Colposcope” section of this manual.
14
Setting Up VideoPath Colposcope 1. Connect the Light Source power cord to the power cord receptacle on the back of the Light Source. 2. Plug the remaining end of the power cord into a properly grounded 110 - 120 volt AC outlet. 3. Connect the S-Video cable to the S-Video output located on the back of the Light Source and connect the other end of the S-Video cable into the line C connector Y/C (S-Video) on the back of the monitor.
itor
’ Line C Y/C In
4. Plug the colposcope connector terminal into the connector port of the Light Source. The umbilical plug should fully engage and “click” into place.
5. Activate the Power switch on the front panel. The green Lamp_ Ignition/Standby light will blink for approximately 11 seconds. Standby LED Light
6. Once the green Lamp Ignition/Standby light remains lit, press the Lamp Ignition Switch. The lamps will ignite and the $$he Lamp Ignition/Standby light will go out. The lamps require 11 seconds to warm Brightness up. This length of time will be indicated Indicator by the center LED of the Picture Brightness Indicator, which blinks until the lamps reach full intensity. Once the Picture Brightness Indicator remains constant, the lamps are fully warmed up, and the system is ready for use.
NOTE: If the Light Source is powered up without having the VideoPath Colposcope attached, the lamp ignition Standby LED will remain blinking and lamp ignition cannot occur.
Operation Power Switch With the Light Source power cord connected to a properly grounded outlet, the S-Video cable connected to the Light Source and monitor, and lamps activated, activate the power switch on the handle of the VideoPath Colposcope.
Focus and Zoom Controls Once the power switch has been activated, the VideopPath Colposcope will execute a setup procedure, ending in the low magnification setting, or 4.5x. To obtain coarse focus, position the distal end of the VideoPath Colposcope approximately 300 mm from the target. Adjust the VideoPath Colposcope by moving the stand (for vertical stand), or the arm (for the swing arm stand), until the picture is in focus. Press the + on one of the zoom controls until maximum magnification is obtained, and readjust the position of the stand or the swing arm until the picture is in focus. Fine focus, in either direction, is provided by the focus controls once coarse focus has been achieved. After coarse focus is achieved at maximum magnification, zoom control can be set to a desired lower magnification by pressing either zoom control. Focus will be maintained throughout the entire magnification range once these steps have been completed, if the VideoPath Colposcope or target is not moved. NOTE: The setup (nominal) setting of the focus controls can be recovered by pressing one of the zoom + controls for 4 seconds after maximum magnification has been achieved.
Green Filter Controls Activate the electronic green filter by pressing the green button located on either side of the VideoPath Colposcope. Full activation of the green filter occurs in less than 10 seconds. Deactivated the filter by pressing the green filter button again. The green filter LED indicator flashes during activation and deactivation and remains on after activation is complete.
Illumination Beam Director The illumination beam director can be rotated up to 45” clockwise or counterclockwise to better illuminate the examination area. To rotate the director, grasp the flat sides and twist in either direction as desired. The illumination beam director moves with some resistance. This helps to hold it securely in the desired position. The center position can be recovered by returning the director to the detent position.
Vertical Height Adjustment Ring (Vertical Colposcope Models 88100 & 88110 Only) The height of the vertical stand can be adjusted by rotating the black height adjustment ring, located on the pole of the stand, counterclockwise. Once the ring has been loosened, adjust the pole to the desired height and then tighten the ring again by rotating it clockwise.
Swing Arm Height Adjustment (Swing Arm Colposcope Models 89100 & 89110 Only) Adjust the height of the swing arm stand by loosening the height adjustment knob located on the arm of the stand. The swing arm can then be moved to the desired position. After moving the arm, tighten knob securely by rotating clockwise. The desired drag can be obtained by adjusting this knob either clockwise or counterclockwise. Right or Left Positioning of Swing Arm (Swing Arm Colposcope Models 89100 & 89110 Only) The swing arm of the VideoPath Colposcope can be positioned to the right or left side of the exam table. To alter the position of the swing arm, pull down the stop pin located under the elbow of the swing arm while rotating the swing arm to the desired side. Release the stop pin once the desired position is achieved. Positioning VideoPath Colposcope The VideoPath Colposcope can be angled up and down and left and right. To change the up/down tilt angle, loosen the knob located on the tilt axis. The VideoPath Colposcope can then be tilted as necessary. To change the horizontal angle, loosen the knob located on the support shaft. The drag of either knob can be set by loosening or tightening as necessary.
16
Remote Video Functions (for VideoPath Colposcope Models 88110 & 89110 Only) The handle of the VideoPath Colposcope contains three buttons that control three video functions. These functions, provided by a remote video printer, are Freeze, Print and Toggle. (See Connection Diagram section of this manual for connecting the printer to the VideoPath Colposcope.) The VideoPath Colposcope models 88110 and 89110, when used with the VLX-20 Light Source and compatible video printer, have the ability to capture, store and print images. The function controls on the VideoPath Colposcope interface with the system for convenient control of the Freeze, Copy and Video features. The RS-232 cable must be connected from the VLX-20 Light Source. Pressing the freeze ("F") button stores a new image in printer memory. The frozen image is displayed on the monitor. Pressing the copy ("C") button prints the image displayed on the monitor. If live video is displayed, the image visible when the button was pressed will be printed. Pressing the video ("V") button changes the display from printer memory (frozen image) to live video, or from live video to printer memory (frozen image), depending on the current mode.
Storage of the VideoPath Colposcope Connector Terminal When the VideoPath Colposcope is removed from the Light Source, the colposcope connector terminal should be stored in the receptacle hanger. 1. To properly place the VideoPath Colposcope connector plug in the hanger, remove the plug from the Light Source.
2. Wrap the cord in a loop using the VideoPath Colposcope connector terminal hooks located on the colposcope stand. 3. Place the VideoPath Colposcope connector terminal into the hanger as indicated. L