AHA Diagnostic ECG Electrode Placement Guide Rev A
Guide
2 Pages
Preview
Page 1
AHA Diagnostic ECG Electrode Placement The quality of an ECG is dependent on the preparation and the resistance between the skin and the electrode. To ensure a good quality ECG and minimize the skin/electrode resistance, remember the following points: 1.) Ensure that the patient is warm and relaxed 2.) Shave electrode area before cleaning 3.) Thoroughly clean the area with alcohol 4.) Let dry prior to applying electrode
Resting ECG RA and LA electrodes should be placed on the wrists. RL and LL electrodes should be placed a few inches above the ankle.
Lead
Electrode Placement
V1
Fourth intercostal space at the right border of the sternum
V2
Fourth intercostal space at the left border of the sternum
V3
Midway between locations V2 and V4
V4
At the mid-clavicular line in the fifth intercostal space
V5
At the anterior axillary line on the same horizontal level as V4
V6
At the mid-axillary line on the same horizontal level as V4 and V5
For service or technical support, please call (800) 535-6663. For supplies, please contact your authorized Welch Allyn distributor.
Advancing Frontline Care™ 71300-0000 Rev. A
IEC Diagnostic ECG Electrode Placement The quality of an ECG is dependent on the preparation and the resistance between the skin and the electrode. To ensure a good quality ECG and minimize the skin/electrode resistance, remember the following points: 1.) Ensure that the patient is warm and relaxed 2.) Shave electrode area before cleaning 3.) Thoroughly clean the area with alcohol 4.) Let dry prior to applying electrode
Resting ECG R and L electrodes should be placed on the wrists. N and F electrodes should be placed a few inches above the ankle.
Lead
Electrode Placement
C1
Fourth intercostal space at the right border of the sternum
C2
Fourth intercostal space at the left border of the sternum
C3
Midway between locations C2 and C4
C4
At the mid-clavicular line in the fifth intercostal space
C5
At the anterior axillary line on the same horizontal level as C4
C6
At the mid-axillary line on the same horizontal level as C4 and C5