Image fusion management Addendum to User manual – ROSA This is an addendum to the ROSA User manual, ref ROSA-041-2.5.8-C User manual. It provides complementary information and warning message when performing a fusion of 3D exams.
Image fusion The image fusion is proposed to the user when a second 3D exam is loaded from the Exam Manager window.
User needs to indicate if the second exam has been acquired in the same coordinates system or if a merge is needed.
When selecting Merge option, user will have to choose between 3 different modes: manual, semi-automatic and automatic (see below). At the end of each merge mode, the user has to confirm that the images are correctly superimposed to ensure device accuracy (see Fusion adjustment section).
When manual or semi-automatic fusion is selected, a window to define matching points in the two exams is displayed.
__________________________________________________________________________________ Image fusion management (addendum) ROSA-401A-EN Page 1 sur 5
a. Automatic fusion When the user selects the automatic fusion mode, an automatic fusion of the 2 selected 3D exams is performed. At the end of the automatic fusion, the fusion result is displayed in the check fusion window, where the surgeon can verify the result of the fusion and performed adjustments. When the images are correctly superimposed to ensure device accuracy, the surgeon validates the fusion by checking the validation box and click on OK to continue. The result is saved and can be restored later.
__________________________________________________________________________________ Image fusion management (addendum) ROSA-401A-EN Page 2 sur 5
b. Manual and semi-automatic fusion When the user selects the semi-automatic merger mode, the surgeon has to identify at least 3 references points in the 2 selected 3D exams, then the fusion is performed. At the end of the semi-automatic fusion, the fusion result is displayed in the check fusion window, where the surgeon can verify the result of the fusion and performed adjustments. When the images are correctly superimposed to ensure device accuracy, the surgeon validates the fusion by checking the validation box and click on OK to continue. The result is saved and can be restored later. When the user selects the manual merger mode, the surgeon has to identify at least 3 references points in the 2 selected 3D exams, then the check fusion window is displayed. The surgeon will perform all required adjustments and checks until the images are correctly superimposed to ensure device accuracy. The surgeon validates the fusion by checking the validation box and click on OK to continue. The result is saved and can be restored later.
1
2
3
To continue the manual or semi-automatic fusion, identify at least 3 reference points in the two exams.
Component
Function
(1)
Display the first image Display the second image List the points defined on the two images. Identify at least 3 reference points in the two exams.
(2) (3)
Released
Pressed
When pressed, it links the navigation. When released, it makes the navigation in each view independent. Records the point common to the two images. Alters the point selected.
__________________________________________________________________________________ Image fusion management (addendum) ROSA-401A-EN Page 3 sur 5
Component
Function Deletes the point selected from the list. Cancels and returns to the registration selection screen. Confirms and passes to the next stage.
c. Fusion adjustment Once a fusion has been performed, a window is displayed that allows the verification of fusion quality and performing a final fine adjustment.
2
1
4
Component
Function
Part 1: Adjust mode for displaying result of fusion Display in transparency mode
Display in two-color mode __________________________________________________________________________________ Image fusion management (addendum) ROSA-401A-EN Page 4 sur 5
3
Component
Function Display in checkerboard mode Adjust display proportions between 2 images
Part 2: Adjust mode for displaying image 1 (fixed) Allows contrast alterations to be applied to image 1 Displays image 1 only. Is not visible in “checkerboard” display mode. Part 3: Adjust mode for displaying image 2 (moving) Displays image 2 only. Is not visible in “checkerboard” mode.
Allows contrast alterations to be applied to image 2
Allows free movement of image 2 Deactivated
Activated
Allows free turning of image 2 Deactivated
Activated
Part 4: Adjustment of transformation of image 2 (moving) Applies a linear or rotatory micro-movement to image 2
Choice of step of linear movement, mm Choice of step of rotatory movement, degrees
Cancels and returns to previous screen Confirms and returns to image management interface. Do not accept a poor fusion quality. If any doubt in the image quality of the merged exam (e.g. arti-fact, blurred areas, shift or other suspicious display in the ROI), review the fusion and / or select another3D exam.
__________________________________________________________________________________ Image fusion management (addendum) ROSA-401A-EN Page 5 sur 5