INSTRUCTIONS (CAUTION LEAFLET) HD CAMERA HEAD
CH-S200-08-LB 1 Compatible Reprocessing Methods and Chemical Agents The camera head is compatible with several methods of reprocessing. However, not all reprocessing methods are compatible with all camera heads. Reprocessing with incompatible methods can cause equipment damage and/or the equipment may not be sterilized. For appropriate reprocessing methods, see Table 1. Follow the policies at your local institution when choosing which methods listed in Table 1 to employ.
List of compatible methods Camera head
Ultrasonic cleaning Manual cleaning
Alkaline enzymatic detergent Neutral enzymatic detergent
Manual disinfection
Peracetic acid Glutaraldehyde
Drying
Alcohol
Automatic cleaning and disinfection
AER
ETD Double (Peracetic acid) ETD 4 (Peracetic acid) ETD 4 (Glutaraldehyde)
WD (Alkaline detergent, thermal disinfection) Sterilization
Hydrogen perioxide
V-PRO® maX (Non Lumen cycle)*1 STERRAD® NX® with ALL Clear Technology (Standard cycle)*2, *3 STERRAD® NX® (Standard cycle)*2 STERRAD® 100NX® with ALL Clear Technology (Duo cycle)*2, *3 STERRAD®100NX® (Duo cycle)*2 STERRAD® 100S (Short cycle)*2
Steam sterilization (autoclaving) Ethylene oxide gas sterilization compatible
Table 1
not compatible
List of compatible methods
*1 Olympus products have been validated by STERIS Corporation for the microbiological efficacy. As for the sterilizer, contact STERIS Corporation. Also, refer to the “V-PRO® maX” instruction manual. *2 Olympus products have been validated by ASP (Advanced Sterilization Products) for the microbiological efficacy. As for the sterilizer, contact ASP. Also, refer to the “STERRAD® 100S/100NX® /NX®” instruction manual. *3 Olympus products can be processed with or without ALLClear™ Technology. For further details, contact ASP.
WARNING • An insufficiently reprocessed camera head device and accessory may pose an infection control risk to the patients and/or operators who contact them. • All disinfection methods (whether performed manually or by a WD), and all sterilization methods (whether performed by ethylene oxide gas, low temperature gas, steam, STERRAD® or V-RPO®) require thorough cleaning of the instruments before. If the instruments are not adequately cleaned prior to disinfection/sterilization, these processes will be ineffective. Immediately after each patient procedure and before disinfection/sterilization, thoroughly clean the camera head and its accessories.
CAUTION Do not sterilize the camera head with steam. Do not reprocess the camera head in an AER. Otherwise, the camera head will be damaged.
Manufactured by 2951 Ishikawa-cho, Hachioji-shi, Tokyo 192-8507, Japan
The meaning(s) of the symbol(s) are as follows:
Fax: (042)646-2429 Telephone: (042)642-2111
Distributed by (Premises/Goods delivery) Wendenstrasse 14-18, 20097 Hamburg, Germany (Letters) Postfach 10 49 08, 20034 Hamburg, Germany Fax: (040)23773-4656 Telephone: (040)23773-0
Manufacturer
Authorized representative in the European Community
3500 Corporate Parkway, P.O. Box 610, Center Valley, PA 18034-0610, U.S.A. Fax: (484)896-7128 Telephone: (484)896-5000
Translation
Importer (into European Union)
5301 Blue Lagoon Drive, Suite 290 Miami, FL 33126-2097, U.S.A. Fax: (305)261-4421 Telephone: (305)266-2332
KeyMed House, Stock Road, Southend-on-Sea, Essex SS2 5QH, United Kingdom Fax: (01702)465677 Telephone: (01702)616333
A8F, Ping An International Financial Center, No. 1-3, Xinyuan South Road, Chaoyang District, Beijing, 100027 P.R.C. Fax: (86)10-5976-1299 Telephone: (86)10-5819-9000
(Seocho-dong, Majestar City Tower One), 2F & 3F, 12, Seocho-daero 38-gil, Seocho-gu, Seoul, 06655, Republic of Korea Fax: (02)6255-3210 Telephone: 1544-3200
Elektrozavodskaya str. 27 bld.8, 107023 Moscow, Russia Fax: (7)495-926-7072 Telephone: (7)495-926-7077
438B Alexandra Road #03-07/12, Alexandra Technopark Blk B, Singapore 119968 Fax: 6834-2438 Telephone: 6834-0010
3 Acacia Place, Notting Hill, VIC 3168, Australia Fax: (03)9543-1350 Telephone: (03)9265-5400
Date of Issue 2021-02-08
RA6572 01
©2021 OLYMPUS MEDICAL SYSTEMS CORP. All rights reserved. Printed in Japan 20210201 *0000