GF-UCT 180 and UE160-AL5 Ultrasonic Gastroscopescopes PreCleaning Guide V1.0 Dec 2019

4 Pages

Ultrasound Gastrovideoscope Precleaning Guide  For: GF-UCT180, GF-UE160-AL5, TGF-UC180J CAUTION: This guide is only a summary of the steps necessary to properly clean your endoscope. Be sure to follow the detailed steps outlined in the INSTRUCTION MANUAL that was included with your endoscope purchase.  Equipment Needed Prepare the following equipment, and wear appropriate protective equipment: n n n n  Personal protective equipment Clean, lint-free cloth 500 ml containers Detergent solution  n n n n  Clean water Air/water channel cleaning adaptor (MAJ-629) Washing tube (MH-974) 5 cc syringe  PRECLEANING - Immediately After Use:  1 Turn OFF the video processor, light source, and ultrasound system. 2 In a clean 500 ml container, prepare clean water. another clean 500 ml container, prepare a compatible detergent solution at the temperature and concentration 3 Inrecommended by the detergent manufacturer. 4 Remove and discard the balloon if attached.  a lint-free cloth soaked in the detergent solution, wipe down the entire insertion tube, from the boot at 5 Using the control section to the distal end. Gently wipe the ultrasound transducer.  MORE INSTRUCTIONS ON NEXT PAGE