14 Pages

en  INSTRUMENT DESCRIPTION The following instructions are for metal reusable general surgical instruments that are characterized by the following design features (examples): •  Sliding shaft instruments that cannot be taken apart (non-dismountable) sliding shaft instruments  non take-apart design (non-dismountable)  •  Tube-center shaft instruments that cannot be taken apart (non-dismountable) irrigation port non take-apart design (non-dismountable)  tube-center shaft instrument  For additional information or guidance on Cleaning and Sterilization instructions, contact your Teleflex Sales Representative or Teleflex Customer Service. INDICATIONS FOR USE Teleflex instruments are designed to perform a specific function, such as cutting, grasping, clamping, dissecting, probing, retracting, draining, aspirating and suturing. Instruments are for use by, or as directed by, a trained medical professional. Instruments should be used only for the purpose for which they are designed. The proper technique for the use of the instrument is the responsibility of the medical professional. PREPARATION AT POINT OF USE Directly after use, remove coarse contamination from the instrument and keep the instrument moist for transit to the processing site. Prior to cleaning and sterilization do not use any fixing agents or hot water (>104°F (40°C)) since this may lead to the fixation of residue and can interfere with the cleaning process. TRANSPORTATION During the transportation of the instrument to the processing site, store contaminated instrument securely in a closed container to avoid damage to the instrument and/or contamination of the environment. PREPARATION FOR REPROCESSING Where applicable, instruments must be disassembled and/or opened for cleaning and sterilization. Ring handled instruments must be opened so that ratchets are not engaged. Failure to properly clean and sterilize the instrument could lead to contaminated instruments being placed back into circulation.  CLEANING, DISINFECTION AND DRYING INSTRUCTIONS Automatic Processing Instructions Manual Processing Instructions For Use with Alkaline or Enzymatic For Use with Enzymatic Detergents Detergents PRE-CLEANING PRE-CLEANING • Soak the instrument in lukewarm • Soak the instrument in lukewarm tap water for a minimum time of tap water for a minimum time of 5 minutes. 5 minutes. • Flush with a 20ml syringe. • Flush with a 20ml syringe. • Brush under lukewarm tap water • Brush under lukewarm tap water with a soft brush until all visible with a soft brush for approximately residues are removed. 40 seconds until all visible residues • Place the instrument in an are removed. ultrasonic bath at 104°F (40°C) with 0.8% solution of enzymatic cleaner. • Sonicate for 10 minutes (minimum). • Remove the instrument from ultrasonic bath and rinse with lukewarm tap water for 15 seconds. CLEANING CLEANING • Place the opened instrument in a • Place the instrument in an perforated basket on the slide-in ultrasound bath at 104°F (40°C) tray and start the cleaning process. with a 0.8% enzymatic detergent. • Pre-rinse for 2 minutes with cold • Sonicate for 10 minutes (minimum). water. • Remove the instrument and • Empty. flush with lukewarm tap water for • Wash for 5 minutes at 131°F (55°C) approximately 15 seconds while with a 0.5% alkaline detergent simultaneously brushing with a or at 131°F (55°C) with a 0.8% soft-bristled brush. enzymatic detergent. • Dry with a lint-free cloth. • Empty. Note: The preparation of • Neutralize for 3 minutes with cold concentration, temperature and water. application time of the cleaning • Empty. agent must be according to the • Rinse for 2 minutes with cold water. instructions for use provided by the • Empty. detergent manufacturer. Examples Note: The preparation of of detergents used in validations: concentration, temperature and Enzymatic Detergent: Cidezyme/ application time of the cleaning Enzol. agent must be according to the instructions for use provided by the detergent manufacturer. Examples of detergents used in validations: Enzymatic Detergent: Cidezyme/ Enzol; Alkaline Detergent: Neodisher Mediclean. DRYING DRYING Dry the outside of the instruments Dry instruments with a lint-free cloth. using the drying cycle of the cleaning Compressed, filtered air can be used equipment. If necessary, instruments for instruments with lumens. can be dried by hand with a lint-free cloth. Compressed, filtered air can be used for instruments with lumens. DISINFECTION DISINFECTION Teleflex does not provide disinfection Teleflex does not provide disinfection instructions due to the sterility instructions due to the sterility instructions provided. instructions provided.
File Type: PDF
File Size: 1.46 MB
File Name: Teleflex Medical - L05523 - Insert, PTD, Generic, Sterility Reproc - 2020-04 - R02.pdf

By clicking Download you are confirming this is the correct document for your purposes
and that you agree to our Terms & Conditions