W&H
Implantology contra-angle handpiece with push-button
Implantology Contra-Angle Dental Handpiece Instructions for Use
48 Pages
Preview
Page 1
Instructions for Use
Implantology Contra-angle handpiece WI-75 E/KM
Contents Symbols... 4 1. Introduction... 6 2. Safety notes... 9 3. Product description... 13 4. Start-up... 14 Assembly/Removal ...14 Rotary instruments...17 To change the rotary instrument... 18 Test run... 20
5. Hygiene and maintenance...21 General notes...21 Limitations on processing...23 Initial treatment at the point of use... 24 Manual cleaning... 25
2
Manual disinfection... 30 Automated cleaning and disinfection... 31 Drying... 32 Inspection, Maintenance and Testing... 33 Packaging... 36 Sterilization... 37 Storage... 40
6. Servicing... 41 7. Accessories, consumables, spare parts and other recommended medical devices by W&H... 42 8. Technical data... 43 9. Disposal... 45 Explanation of warranty terms... 46 Authorized W&H service partners... 47
3
Symbols
WARNING! (risk of injury)
ATTENTION! (to prevent damage occurring)
General explanations, without risk to persons or objects
Caution! According to Federal law, this medical device may only be sold by or on the order of a dentist, physician or any other medical practitioner licensed by the law of the State in which he or she practices and intends to use or order the use of this medical device.
4
Do not dispose of with domestic waste
Manufacturer
Symbols XXXX
CE marking with identification number of the Notified Body
DataMatrix Code for product information including UDI (Unique Device Identification)
Data structure in accordance with Health Industry Bar Code
Catalogue number
Thermo washer disinfectable
Sterilizable up to the stated temperature
Serial number
UL Component Recognition Mark indicates compliance with Canadian and U.S. requirements
Medical device
Date of manufacture
Consult Instructions for Use
5
1. Introduction Customer satisfaction has absolute priority in the W&H quality policy. This medical device has been developed, manufactured and subjected to final inspection according to legal regulations, quality and industry standards. For your safety and the safety of your patients Prior to initial use please read the Instructions for Use. These explain how to use your medical device and guarantee a smooth and efficient operation. Observe the safety notes. Intended use Surgical treatment of organic hard tissue. Intended part of the body > Mouth > Mouth Space 6
Misuse may damage the medical device and hence cause risks and hazards for patient, user and third parties. Qualifications of the user We have based our development and design of the medical device on the physician target group. Responsibility of the manufacturer The manufacturer can only accept responsibility for the safety, reliability and performance of the medical device when it is used in compliance with the following directions: > The medical device must be used in accordance with these Instructions for Use. > The medical device has no components that can be repaired by the user. > Modifications or repairs must only be undertaken by an authorized W&H service partner (see page 47).
7
Skilled application The medical device is intended only for skilled application according to the intended use as well as in compliance with the valid health and safety at work regulations, the valid accident prevention regulations and in compliance with these Instructions for Use. The medical device should be prepared for use and maintained by staff who have been trained in procedures for infection control, personal safety and patient safety. Improper use, (e.g., through poor hygiene and maintenance), non-compliance with our instructions or the use of accessories and spare parts which are not approved by W&H, invalidates all claims under warranty and any other claims. Any serious incident that has occurred in relation to the medical device should be reported to the manufacturer and the competent authority!
8
2. Safety notes > Before using the medical device for the first time, store it at room temperature for 24 hours. > The operation of the medical device is permitted only on supply units which correspond to the standards IEC 60601-1 (EN 60601-1) and IEC 60601-1-2 (EN 60601-1-2). > Always ensure the correct operating conditions and cooling function. > Always ensure that sufficient and adequate cooling is delivered and ensure adequate suction. > In case of coolant supply failure, the medical device must be stopped immediately. > Check the medical device for damage and loose parts each time before using. > Do not operate the medical device if it is damaged. > Only attach the medical device onto the motor when the motor is at a complete standstill.
9
> Never touch the rotary instrument while it is still rotating. > Perform a test run each time before using. > Avoid overheating at the treatment site. > Do not touch the soft tissue with the handpiece head (risk of burning due to the push-button heating up)!
10
The medical device has a considerably higher level of efficiency than normal contra-angle handpieces and has been designed to fit W&H drive units. When using the medical device on other surgical units, the user assumes sole responsibility. The manufacturer accepts no liability. Any concessions can be granted directly by the manufacturer of the drive units.
11
Hygiene and maintenance prior to initial use > The medical device is sealed in PE film and not sterilized when delivered. > The PE film and the packaging are non-sterilizable. > Clean, disinfect and lubricate the medical device. > Sterilize the medical device, the nozzle cleaner, the Y-manifold, the spray clips and the coolant tube for internal bur cooling system.
12
3. Product description Push-button Hexagon chucking system Coolant tube for internal bur cooling Spray clip for Kirschner/Meyer internal bur cooling system
Spray clip without internal bur cooling system
Y-manifold Nozzle cleaner
13
4. Start-up
Assembly/Removal Spray clip Coolant tube for internal bur cooling system
Attach the spray clip. Insert the coolant tube for internal bur cooling system.
Check the secure hold of the spray clip and replace it if necessary or
Remove the coolant tube for internal bur cooling system.
Remove the spray clip. 14
Medical device Do not assemble or remove the medical device during operation!
Push the medical device onto the motor and turn until it engages audibly.
When using a locking pin between motor and medical device: See Instructions for Use of the control unit.
Verify full engagement. or
Remove the medical device.
15
Y-manifold Only when using the coolant tube for internal bur cooling system.
Insert the coolant hose of the Y-manifold onto the
coolant tube of the spray clip and onto the coolant tube for internal bur cooling system. Insert the Y-manifold into the irrigation tubing. or
Remove the Y-manifold.
16
Rotary instruments > Use only rotary instruments which are in perfect condition and pay attention to the direction of rotation of the rotary instrument. Follow the operating instructions of the manufacturer. > Insert the rotary instrument only when the medical device is stationary. > Never touch the rotary instrument while it is still rotating. > Do not activate the push-button of the medical device during operation. This leads to detachment of the rotary instrument, damage to the chucking system and/or heating up of the medical device. Risk of burning! When having a torque higher than 30 Ncm on the rotary instrument you have to use hardened shafts (>50 HRC, >520 HV) (risk of deformation).
17
To change the rotary instrument > Using internal bur cooling system > Without using internal bur cooling system > With hexagon > Instrument shaft diameter 2.35 mm
Insert the coolant tube for internal bur cooling system.
or
Remove the coolant tube for internal bur cooling system.
Insert the rotary instrument over the coolant tube until the limit stop (a).
Activate the push-button and turn the rotary instrument until it engages (b).
18
Verify full engagement. or
Activate the push-button and remove the rotary instrument.
Ask your implant supplier for an applicable rotary instrument with hexagon.
19
Test run Do not hold the medical device at eye level! > Insert the rotary instrument. > Operate the medical device. In the event of operating malfunctions (e.g., vibrations, unusual noise, overheating, coolant failure or leakage) stop the medical device immediately and contact an authorized W&H service partner.